CL17 Analyser for total chlorine

NIEAKTUALNA POZYCJA

CL17 Analyser for total chlorine
Nr produktu: 5440002

NIEAKTUALNA POZYCJA

Colorimetric chemistry - simple, fast, accurate

Fast automatic operation

Complete control - minimal maintenance

Rugged, lightweight enclosure

Zawartość opakowania

CL17 Chlorine Analyser with 1 Month Supply of Total Chlorine Reagents, Maintenance kit, Installation Kit, and Manual. (The power cord is ordered separately.)

Parametry techniczne

Akcesoria: Żaden
  Żaden
Alarmy: Two alarms selectable for sample concentration alarm, analyser system warning, or analyser system shut-down alarm. Each is equipped with an SPDT relay with contacts rated for 5A resistive load at 230 V AC.
Ciśnienie na wlocie: 1 to 5 psi (0.07 - 0.34 bar), 1.5 psi (0.103 bar) is optimum
Czas cyklu: 2.5 min
Czyszczenie powietrzem: 0.1 cfm (0.17 m³/h) instrument quality air at max. 20 psi (ca.1.4 bar) with ¼" OD tube, quick disconnect fitting
Dokładność: ± 5 % or ±0.03 mg/L (ppm) as CL2, whichever is greater
Gwarancja: 24 miesięcy
Limit detekcji: 0.03 mg/L (ppm)
Obudowa: ABS plastic, two clear polycarbonate windows, IP62-rated with the gasketed door latched.
Opcje komunikacji: 4-20 mA output
Opis: CL17 Chlorine Analyzer with Reagents for Total Chlorine
Parametr: Chlorine, Free & Total
Powtarzalność: 0.01 ppm lub 3.0 % , którakolwiek jest większa
Prędkość przepływu: 200 - 500 mL na minutę (wymagane minimum)
Pressure Regulator Included?: No
Sposób montażu: Wall mount
Temperatura próbki: 5 - 40 °C (41 - 104 °F)
Waga: 11,3 kg
Wilgotność: Up to 90 % at 40 °C (104 °F) maximum
Wlot: ¼" OD polyethylene tube, quick-disconnect fitting
Wyjścia rejestratora: Jedno 0/4...20 mA dowolnie programowalne w zakresie od 0 do 5 mg/L. Zalecana impedancja obciążenia od 3,6 do 500 omów, uziemienie 130 V.
Wymogi energetyczne: 2,5 A
Wymogi energetyczne (Hz): 50/60 Hz
Wymogi energetyczne (napięcie): 100 - 115/230 VAC
Wyświetlacz: LCD, 3.5 inch digit measurement readout and six-character alphanumeric scrolling text line
Zakłócenia: Other oxidizing agents such as bromide, chlorine dioxide, permanganate and ozone will cause a positive interference. Hexavalent chromium will cause a positive interference:
1 mg/L Cr6+ = approximately 0.02 mg/L as Cl2.
Hardness must not exceed 1,000 mg/L CaCO3.
Zakres ciśnienia: Inlet Pressure to Sample Conditioning: 1.5 - 75 psi (0.1 - 5.2 bar)
Zakres pomiarowy: 0 - 5 mg/L Cl₂
Zakres temperatury pracy: 5 - 40 °C
Zawartość opakowania: CL17 Chlorine Analyser with 1 Month Supply of Total Chlorine Reagents, Maintenance kit, Installation Kit, and Manual. (The power cord is ordered separately.)
  CL17 Chlorine Analyser with 1 Month Supply of Total Chlorine Reagents, Maintenance kit, Installation Kit, and Manual. (The power cord is ordered separately.)
Źródło światła: Class 1 LED (light emitting diode) with a peak wavelength of 520 nm; 50,000 hours estimated minimum life