Akcesoria:
Żaden
Żaden
Alarmy:
Two alarms selectable for sample concentration alarm, analyzer system warning, or analyzer system shut-down alarm. Each is equipped with an SPDT relay with contacts rated for 5A resistive load at 230 V AC.
Certyfikaty:
Europe, CE Approved with:
EN 61326-1
CISPR 11
EN 61010-1
IEC 60529
North America:
UL 61010A-1
CAN/CSA C22.2 No. 1010.1-92
Ciśnienie na wlocie:
1 - 5 psig (0.07 to 0.34 bar), 1.5 psi is optimum
Czas cyklu:
2.5 min
Czyszczenie powietrzem:
0.1 cfm (0.17 m³/h) instrument quality air at max. 20 psig (ca. 1.4 bar) with 1/4-inch OD tube, quick disconnect fitting
Dokładność:
± 5 % or ±0.03 mg/L (ppm) as CL₂, whichever is greater. *For permanganate accuracy, please refer to permanganate application notes
Gwarancja:
24 miesięcy
Limit detekcji:
0.03 mg/L (ppm)
Obudowa:
ABS plastic, two clear polycarbonate windows, IP62-rated with the gasketed door latched.
Odpływ:
1/2-inch ID flexible hose, hose barb
Opcje komunikacji:
4-20 mA output
Opis:
CL17 Chlorine Analyzer (Instrument Only)
Parametr:
Chlorine, Free & Total
Powtarzalność:
0.01 ppm lub 3.0 % , którakolwiek jest większa
Prędkość przepływu:
200 - 500 mL na minutę (wymagane minimum)
Pressure Regulator Included?:
No
Sposób montażu:
Wall mount
Temperatura próbki:
5 - 40 °C (41 - 104 °F)
Waga:
10,49 kg
Wilgotność:
Up to 90 % at 40 °C (104 °F) maximum
Wlot:
1/4-inch OD polyethylene tube, quick-disconnect fitting
Wyjścia rejestratora:
Jedno 0/4...20 mA dowolnie programowalne w zakresie od 0 do 5 mg/L. Zalecana impedancja obciążenia od 3,6 do 500 omów, uziemienie 130 V.
Wyświetlacz:
LCD, 3-1/2 inch digit measurement readout and six-character alphanumeric scrolling text line
Zakłócenia:
Other oxidizing agents such as bromide, chlorine dioxide, permanganate and ozone will cause a positive interference. Hexavalent chromium will cause a positive interference: 1 mg/L Cr⁶⁺ = approximately 0.02 mg/L as Cl₂. Hardness must not exceed 1,000 mg/L as CaCO₃.
Zakres ciśnienia:
Inlet Pressure to Sample Conditioning: 1.5 - 75 psi (0.1 - 5.2 bar)
Zakres pomiarowy:
0 - 5 mg/L Cl₂
Zakres temperatury pracy:
5 - 40 °C (41 - 104 °F)
Zawartość opakowania:
CL17 Chlorine Analyser, Maintenance kit, Installation Kit, and Manual. (The power cord is ordered separately.)
CL17 Chlorine Analyser, Maintenance kit, Installation Kit, and Manual. (The power cord is ordered separately.)
Źródło światła:
Class 1 LED (light emitting diode) with a peak wavelength of 520 nm; 50,000 hours estimated minimum life