Alarmy:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Certyfikaty:
CE compliant / UL certified
Ciśnienie próbki:
Maximum 3 bar for direct injection or by external overflow vessel via the grab sample pump
Czas cyklu:
7 - 10 minutes incl. sample lysis
Demineralised water:
For rinsing
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Gwarancja:
24 miesięcy
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Kalibracja:
Automatic, 2-point; frequency freely programmable
Klasa ochrony:
Analyser cabinet: IP44 / Panel PC: IP65
Liczba strumieni próbki:
1, 2, 4, or 8
Limit detekcji:
≤ 0.05 pg/mL (0.1 pM) ATP
Materiał:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metoda pomiarowa:
Determination of adenosine triphosphate (ATP) by means of chemiluminescent reaction using luciferin and luciferase, conform with standard method ASTM D4012-81
Model:
EZ7300
Odpływ:
Atmospheric pressure, vented, min. Ø 32 mm
Parametr:
ATP
Pobór mocy:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Precyzja:
Better than 4% full scale range for standard test solutions
Prędkość przepływu:
100 - 300 mL/min
Reagent Requirements:
Keep between 4 - 8 °C during operation, shielded from light. For prolonged storage, store at -20 °C in a dark environment.
Temperatura otoczenia:
7 - 23 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura próbki:
5 - 30 °C
Validation:
Automatic; frequency freely programmable
Waga:
30 kg
Wyjścia analogowe:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Wyjścia cyfrowe:
Optional: Modbus (TCP/IP, RS485)
Wymiary (wys. x szer. x głęb.):
690 mm x 465 mm x 330 mm
Zakłócenia:
High concentrations of Hg2+, Cu2+, Zn2+, Cd2+, Fe2+. Total salt concentrations higher than 1 g/L. pH lower than 5.5 and higher than 8.
Zakres pomiarowy:
0.5 - 200 pg/mL
0.5 - 200 pg/mL
Zasilanie:
100 - 240 VAC, 50/60 Hz
Zawartość opakowania:
EZ7300 ATP Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 2 x empty 5L Reagent Container with Fittings (NaOH & HCl Solution) and 3 x Screwcap/Fittings/Reference1/Reference2