Alarmy:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Automatyczne czyszczenie:
Tak
Certyfikaty:
CE compliant / ETL certified
Ciśnienie próbki:
By external overflow vessel
Czas cyklu:
15 min (dilution + 5 min.)
Demineralised water:
When needed for rinsing / dilution
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Gwarancja:
24 miesięcy
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Jakość próbki:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Kalibracja:
Automatic, 2-point; frequency freely programmable
Klasa ochrony:
IP44
Liczba strumieni próbki:
Standard: 1
Optional accessory for 2, 4, or 8 sample streams
Limit detekcji:
≤ 50 µg/L
Materiał:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metoda pomiarowa:
Colorimetric measurement at 670 nm using N,N-dimethyl-p-phenylenediamine dihydrochloride (NNDP), conform with standard method ASTM 4500-S2- D
Odpływ:
Atmospheric pressure, vented, min. Ø 32 mm
Parametr:
Sulphide
Pobór mocy:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Precyzja:
Better than 2% full scale range for standard test solutions
Prędkość przepływu:
100 - 300 mL/min
Reagent Requirements:
Keep between 5 - 20 °C in use
Temperatura otoczenia:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura próbki:
10 - 30 °C
Validation:
Automatic; frequency freely programmable
Waga:
40 kg
Woda chłodząca:
Flow rate approx. 5 L/h; temperature max. 30 °C; pressure max. 0.5 bar
Wyjścia analogowe:
Active 4 - 20 mA max. 500 Ohm load, standard 4, max. 8 (option)
Wyjścia cyfrowe:
None
Optional: Modbus (TCP/IP, RTU), Profinet, Profibus DP, Ethernet IP
Wymiary (wys. x szer. x głęb.):
688 mm x 460 mm x 340 mm
Zakłócenia:
Strong reducing agents like Hydrosulphide, Nitrite > 0,2 mg/L, Sulphite > 200 mg/L and Thiosulphate > 5 mg/L interfere by reducing the blue colour. Oxygen reacts with Sulphide and must be avoided. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Zakres pomiarowy:
0.025 - 0.5 mg/L
0.125 - 2.5 mg/L (with internal dilution)
0.25 - 5 mg/L (with internal dilution)
0.625 - 12.5 mg/L (with internal dilution)
1.25 - 25 mg/L (with internal dilution)
1.875 - 37.5 mg/L (with internal dilution)
2.5 - 50 mg/L (with internal dilution
Zawartość opakowania:
EZ1037sc Sulphide Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and 2 x empty 2.5L Reagent Glass Containers with Fittings (for Colour, Buffer)