Alarmy:
1x malfunctioning, 4x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Certyfikaty:
CE compliant / UL certified
Ciśnienie próbki:
By external overflow vessel
Czas cyklu:
10 - 15 minutes
Demineralised water:
For rinsing / dilution
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Gwarancja:
24 miesięcy
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Jakość próbki:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Kalibracja:
Automatic, 2-point; frequency freely programmable
Klasa ochrony:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Liczba strumieni próbki:
1, 2, 4 or 8
Limit detekcji:
≤ 10 mg/L
Materiał:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metoda pomiarowa:
Acid-base titration with hydrochloric acid
Nazwa metody:
Titrimetric
Odpływ:
Atmospheric pressure, vented, min. Ø 32 mm
Parametr:
Alkalinity, free
Pobór mocy:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Precyzja:
Better than 2% full scale range for standard test solutions
Prędkość przepływu:
100 - 300 mL/min
Reagent Requirements:
Keep between 10 - 30 °C
Temperatura otoczenia:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura próbki:
10 - 30 °C
Validation:
Automatic; frequency freely programmable
Waga:
25 kg
Wyjścia analogowe:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Wyjścia cyfrowe:
Optional: Modbus (TCP/IP, RS485)
Wymiary (wys. x szer. x głęb.):
690 mm x 465 mm x 330 mm
Zakłócenia:
Soaps, oily matter, suspended solids or precipitates may coat the glass electrode and cause a sluggish response. Allow additional time between titrant additions to let the electrode come to equilibrium or clean the electrodes occasionally.
Zakres pomiarowy:
100 - 5000 mg/L CaCO3
Optional:
10 - 500 mg/L CaCO3
25 - 1250 mg/L CaCO3
50 - 2500 mg/L CaCO3
Internal dispenser dilution (max. factor 100)
100 - 5000 mg/L CaCO₃