Alarmy:
                                            
                                            
                                        
									    1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
                                    
                                
    						
								
                                    
                                        
                                            
                                                Certyfikaty:
                                            
                                            
                                        
									    CE compliant / ETL certified
                                    
                                
    						
								
                                    
                                        
                                            
                                                Ciśnienie próbki:
                                            
                                            
                                        
									    By external overflow vessel
                                    
                                
    						
								
                                    
                                        
                                            
                                                Czas cyklu:
                                            
                                            
                                        
									    10 - 15 minutes
                                    
                                
    						
								
                                    
                                        
                                            
                                                Demineralised water:
                                            
                                            
                                        
									    N.A.
                                    
                                
    						
								
                                    
                                        
                                            
                                                Earth connection:
                                            
                                            
                                        
									    Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
                                    
                                
    						
								
                                    
                                        
                                            
                                                Gwarancja:
                                            
                                            
                                        
									    24 miesięcy
                                    
                                
    						
								
                                    
                                        
                                            
                                                Instrument air:
                                            
                                            
                                        
									    Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
                                    
                                
    						
								
                                    
                                        
                                            
                                                Jakość próbki:
                                            
                                            
                                        
									    Maximum particle size 500 µm, < 0.1 g/L
Most applications require the use of an EZ9130 sampling/filtration system.
                                    
                                
    						
								
                                    
                                        
                                            
                                                Kalibracja:
                                            
                                            
                                        
									    Automatic; frequency freely programmable
                                    
                                
    						
								
                                    
                                        
                                            
                                                Klasa ochrony:
                                            
                                            
                                        
									    Analyser cabinet: IP44 / Panel PC: IP65
                                    
                                
    						
								
                                    
                                        
                                            
                                                Liczba strumieni próbki:
                                            
                                            
                                        
									    1, 2, or 4
                                    
                                
    						
								
                                    
                                        
                                            
                                                Limit detekcji:
                                            
                                            
                                        
									    ≤ 10 mg/L
                                    
                                
    						
								
                                    
                                        
                                            
                                                Materiał:
                                            
                                            
                                        
									    Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
                                    
                                
    						
								
                                    
                                        
                                            
                                                Metoda pomiarowa:
                                            
                                            
                                        
									    Acid-base titration
                                    
                                
    						
								
                                    
                                        
                                            
                                                Nazwa metody:
                                            
                                            
                                        
									    Titrimetric
                                    
                                
    						
								
                                    
                                        
                                            
                                                Odpływ:
                                            
                                            
                                        
									    Atmospheric pressure, vented, min. Ø 32 mm
                                    
                                
    						
								
                                    
                                        
                                            
                                                Parametr:
                                            
                                            
                                        
									    Volatile fatty acids (VFAs), total alkalinity
                                    
                                
    						
								
                                    
                                        
                                            
                                                Pobór mocy:
                                            
                                            
                                        
									    100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
                                    
                                
    						
								
                                    
                                        
                                            
                                                Precyzja:
                                            
                                            
                                        
									    Better than 3% full scale range for standard test solutions
                                    
                                
    						
								
                                    
                                        
                                            
                                                Prędkość przepływu:
                                            
                                            
                                        
									    100 - 300 mL/min
                                    
                                
    						
								
                                    
                                        
                                            
                                                Reagent Requirements:
                                            
                                            
                                        
									    Keep between 10 - 30 °C
                                    
                                
    						
								
                                    
                                        
                                            
                                                Rinsing:
                                            
                                            
                                        
									    With tap water
                                    
                                
    						
								
                                    
                                        
                                            
                                                Temperatura otoczenia:
                                            
                                            
                                        
									    10 - 30 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Temperatura próbki:
                                            
                                            
                                        
									    10 - 30 °C
                                    
                                
    						
								
                                    
                                        
                                            
                                                Validation:
                                            
                                            
                                        
									    Automatic; frequency freely programmable
                                    
                                
    						
								
                                    
                                        
                                            
                                                Waga:
                                            
                                            
                                        
									    25 kg
                                    
                                
    						
								
                                    
                                        
                                            
                                                Wyjścia analogowe:
                                            
                                            
                                        
									    Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Wyjścia cyfrowe:
                                            
                                            
                                        
									    Optional: Modbus (TCP/IP, RS485)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Wymiary (wys. x szer. x głęb.):
                                            
                                            
                                        
									    690 mm x 465 mm x 330 mm
                                    
                                
    						
								
                                    
                                        
                                            
                                                Zakłócenia:
                                            
                                            
                                        
									    Phosphates and similar dissociating ions and non-fatty acids which on acidification from undissociated acids may cause interference. Sulphide may deteriorate some types of pH electrodes. Fats, oil, proteins, surfactants and tar.
                                    
                                
    						
								
                                    
                                        
                                            
                                                Zakres pomiarowy:
                                            
                                            
                                        
									    VFA: 100‐5000 mg/L, Total Alk: 250 - 5000 mg/L
                                    
                                
                                    
                                        
                                            
                                            
                                                 
                                            
                                        
									    VFAs 100 - 5000 mg/L as acetate equivalent; 
Alkalinity, total and Alkalinity, partial 250 - 5000 mg/L as CaCO₃
                                    
                                
    						
								
                                    
                                        
                                            
                                                Zawartość opakowania:
                                            
                                            
                                        
									    EZ7252 VFA & Alkalinity Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x pH Electrode, Routine S7 Connection, 2 x empty 10L Reagent Containers with Fittings (NaOH & H₂SO₄ Solution) and 1 x empty 2.5L Reagent Container with Fittings (Anti-foam Solution)