Alarmy:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Automatyczne czyszczenie:
Tak
Certyfikaty:
CE compliant / ETL certified
Ciśnienie próbki:
By external overflow vessel
Czas cyklu:
Standard measurement cycle time for both TN & TP: 60 minutes
Demineralised water:
Consumption: 140 mL/analysis (hour)
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Gwarancja:
24 miesięcy
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Jakość próbki:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Kalibracja:
Automatic, 2-point; frequency freely programmable
Klasa ochrony:
Analyser cabinet: IP55 / Panel PC: IP65
Liczba strumieni próbki:
1, 2, 4, or 8
Limit detekcji:
TN: ≤ 0.5 mg/L
TP: ≤ 0.01 mg/L
Materiał:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metoda pomiarowa:
TN: UV photometric measurement at 220 nm after persulphate digestion in alkaline medium, based on APHA 4500-NO3 (B)
TP: Colorimetric measurement at 700 nm using ascorbic acid reduction and molybdate colour solution after persulphate digestion in acidic medium, based on APHA 4500-P
Odpływ:
Atmospheric pressure, vented, min. Ø 32 mm
Parametr:
Total Nitrogen (TN), Total Phosphorus (TP)
Pobór mocy:
230 VAC, 50/60 Hz
120 VAC, 50/60 Hz
Max. power consumption: 440 VA
Precyzja:
Better than 3% (TN) and 2% (TP) full scale range for standard test solutions
Prędkość przepływu:
100 - 300 mL/min
Reagent Requirements:
Keep between 10 - 30 °C
Temperatura otoczenia:
10 - 30 °C, ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura próbki:
10 - 30 °C
Validation:
Automatic; frequency freely programmable
Waga:
25 kg
Wyjścia analogowe:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Wyjścia cyfrowe:
Optional: Modbus (TCP/IP, RS485)
Wymiary (wys. x szer. x głęb.):
690 mm x 465 mm x 330 mm
Zakłócenia:
TN: The main interferences are Br- and I-. When the amount of I- is 2.2 fold of the amount of TN, or the amount of Br- is 3.4 fold of the amount of TN, this will interfere on the test results. Dissolved organic matter, surfactants and Chromium (VI) interfere. Various inorganic substances not normally found in natural water, such as Chlorite [ClO₂-] and Chlorate [ClO₃-], may interfere.
TP: Arsenic (V), Chromium (VI), Copper (II) > 10 mg/L, Iron (III) > 10 mg/L, Sulphide > 2 mg/L, and Vanadium (V), Silica > 60 mg/L. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Zakres pomiarowy:
0.5 - 20 mg/L TN; 0.010 - 2 mg/L TP
Zawartość opakowania:
EZ7652 Total Nitrogen & Total Phosphate Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 3 x empty 2.5L Reagent Containers with Fittings (HCl, NaOH and H₂SO₄ Solution) and 3 x empty 2.5L Glass Reagent Containers with Fittings (Colour, Reductor & Persulphate Solution)